Why Vacuum Pro?

All-in-One: for the sterilization of solids and liquids

Safe: reliable sterilization thanks to integrated media sensor

Fast: short sterilization cycles from 20 minutes

Programmable: saving time through standard programs and automatic repeat function

Reliable: robust technology, high-quality components from brand manufacturers, vacuum pump Made in Germany

Documentation: automatic documentation via CertoCloud, GLP-conform


Technical features

Performance: sterilization temperature 105 ° C to 134 ° C, maximum operating pressure 230 kPa

Method: vacuum, pressure, steam

Volume: 12 / 18 / 22 liter

Control: via the 7" touchscreen or the CertoCloud web interface

Water supply: integrated water tank / by water connection (with RO system)

Programs: 10 standard programs for solids and liquids, all parameters editable

Automation: hundreds of sterilization cycles in a row - fully automatic - without user interaction, ideal for material testing


Programmable

  • select from a number of predefined standard programs, all parameters can be adjusted for optimal sterilization
  • the automatic repeat function allows material testing with any number of sterilization cycles without user interaction. With the included CertoCloud software every single cycle can (even remotely) be monitored in real time and logged. The notification function informs the user via SMS or email when the process has completed.


7'' Touchscreen

  • user management
  • program selection and start
  • program editor
  • information and training portal
  • log viewing
  • integrated step-by-step-video-assistant

CertoCloud - autoclave management software

·         access data from your autoclave in real-time, even remotely

·         configure SMS and E-Mail notifications

·         view protocols and export them to signed PDF, Excel or ZIP

·         confirm the authenticity and validity of printed out protocols with a few clicks

 


Sterilization of solids

  • all sterilization programs for solids follow a typical Class-B cycle
  • precise combination of vacuum, pressure and temperature for optimal sterilization of solids
  • the fractionated vacuum presses the saturated water vapor through the sterilization package and into each microscopic gap


Sterilization of liquids

  • all sterilization programs for liquids follow a heating phase with venting followed by a sterilization phase and a final cooling phase
  • sterilize liquids at 121°C for 20 minutes
  • the integrated media sensor measures the real temperature of liquids and ensures an optimum sterilization result

* suggested retail price. Only applicable in Germany and Austria